Preclinical Imaging

M3™ System Specifications

The M3™ compact, high performance MRI system is designed to generate high resolution 3D anatomical, functional and molecular images with a bore size optimized for mice.

The system’s small footprint and lightweight design enables flexible placement where its needed most even behind the bio containment barrier for in vivo imaging.

M3-no-background

Gradients

Strength 440 (X,Y), 540 (Z) mT/m @ 60 A
Linearity
(for Entire Imaging Volume)
Better than 4%
Slew Rate (T/m/s) 1750 T/m/s @ 60 A / 250 us

Imaging Pulse Sequences

Sequence Details
Spin Echo 2D
DWI – with diffusion gradients
IR – with preceding inversion pulse
Fast Spin Echo 2D and 3D
DWI – with diffusion gradients
IR – with preceding inversion pulse
Gradient Echo 2D and 3D
IRSnap GRE based with multiple inversion time
DCE GRE based Dynamic Contrast Enhancement
Cardiac Cine GRE based ECG triggered Cine
DTI FSE based diffusion tensor imaging
Dixon FSE based fat separation
Field of View 80 x 80 x 35 mm³
spheroid

Back End — Electronics Cabinet

Dimensions (Height x Width x Depth) 1760 x 600 x 1150 mm / 69.3 x 23.6 x 45.3 in
Weight 300 Kg / 660 Lbs

Front End — Magnet & Stand

Dimensions
(Height x Width x Depth)
1080 x 734 x 734 mm /
42.5 x 29 x 29 in
Weight 560 Kg / 1232 Lbs
Magnet Opening:
Flange Insertion Diameter
Inner Bore
(Height x Width)
70 mm / 2.8 in
50 x 130 mm / 2 x 5.1 in
Imaging Volume 80 x 80 x 35 mm³
spheroid
B° (Tesla) 1.0 T

Electrical Requirements

Operator Station Requires an electrical outlet with the following specifications: 120/230 VAC, single phase, 3 A 50/60 Hz
Line Voltage • 220-240 VAC, 15 A, single phase in countries using 220-240V
• 208 VAC, 15 A, single phase in countries using 120V
M3-Spec-Sheet-Cover-Img-

System Specifications and Accessories

Click below to download the full product specifications for the
M3™ Compact MRI system.

Get The Answers You Need.

Have questions? Speak with one of our specialists to learn how the M Series can deliver quality imaging to support your research needs.

Aspect Imaging
Privacy Overview

This website uses cookies so that we can provide you with the best user experience possible. Cookie information is stored in your browser and performs functions such as recognizing you when you return to our website and helping our team to understand which sections of the website you find most interesting and useful.

To view our full privacy policy, please visit this link: https://aspectimaging.guybenami.com/privacy-policy/