The M3™ compact, high performance MRI system is designed to generate high resolution 3D anatomical, functional and molecular images with a bore size optimized for mice.
The system’s small footprint and lightweight design enables flexible placement where its needed most even behind the bio containment barrier for in vivo imaging.
| Strength | 440 (X,Y), 540 (Z) mT/m @ 60 A |
|
Linearity (for Entire Imaging Volume) |
Better than 4% |
| Slew Rate (T/m/s) | 1750 T/m/s @ 60 A / 250 us |
| Sequence | Details |
|---|---|
| Spin Echo | 2D DWI – with diffusion gradients IR – with preceding inversion pulse |
| Fast Spin Echo | 2D and 3D DWI – with diffusion gradients IR – with preceding inversion pulse |
| Gradient Echo | 2D and 3D |
| IRSnap | GRE based with multiple inversion time |
| DCE | GRE based Dynamic Contrast Enhancement |
| Cardiac Cine | GRE based ECG triggered Cine |
| DTI | FSE based diffusion tensor imaging |
| Dixon | FSE based fat separation |
| Field of View | 80 x 80 x 35 mm³ spheroid |
| Dimensions (Height x Width x Depth) | 1760 x 600 x 1150 mm / 69.3 x 23.6 x 45.3 in |
| Weight | 300 Kg / 660 Lbs |
|
Dimensions (Height x Width x Depth) |
1080 x 734 x 734 mm / 42.5 x 29 x 29 in |
| Weight | 560 Kg / 1232 Lbs |
|
Magnet Opening: Flange Insertion Diameter Inner Bore (Height x Width) |
70 mm / 2.8 in 50 x 130 mm / 2 x 5.1 in |
| Imaging Volume | 80 x 80 x 35 mm³ spheroid |
| B° (Tesla) | 1.0 T |
| Operator Station | Requires an electrical outlet with the following specifications: 120/230 VAC, single phase, 3 A 50/60 Hz |
| Line Voltage |
• 220-240 VAC, 15 A, single phase in countries using 220-240V • 208 VAC, 15 A, single phase in countries using 120V |